Web Interface | 5-7
Trends Menu
Click the Trends button to access parameters (Figure 5-5) for:
Sample Cell Temperature
Sample Cell Pressure
Analyte Concentration
Reference Temperature
Laser Temperature
Trend intervals for parameters are available in 30-second, 3005-seconds or
1800-second data points.
The analyzer stores 3600 parameter data points in temporary
memory before overwriting the data. At 30-second intervals this is
1800 minutes or 30 hours; and at 1800 seconds this is 1800 hours
(75 days).
The Trends window displays the most recent data held in short-term memory at
the upper-right edge of the Trend chart. Since data cannot be saved, use the Print
Screen key or other sceen capture methods to copy and paste data screens for
troubleshooting.
Click the Get Trend button to refresh the display with the most
recent data, after changing the interval or selecting a dierent
parameter.
NOTE
NOTE