User Interface | 4-21
Data Type Description MB Starting
Regs
MB Reg Size
char[20] Serial Number 0 10
char[6] MAC address 10 3
uint IP Address 13 2
uint IP Subnet Mask 15 2
uint Gateway 17 2
uint DNS 19 2
uint16 DHCP enable. 1=disabled, 2=enabled 21 1
uint16 TCP Enable. 0=disabled, 1=enabled 22 1
uint Server IP 23 2
int16 Version of the params.xml le generated by
this
25 1
char[20] Software Version 26 10
char[20] Model Name 36 10
uint Internal UART1baud rate (MODBUS) 46 2
uint Internal UART2 baud rate (DISPLAY) 48 2
uint External UART1 baud rate (RS232) 50 2
uint External UART2 baud rate 52 2
uint16 Interface Board FW version 54 1
uint16 0x20: HC121, 0x40: HC122 55 1
int16 Parity. 0=none, 1=even, 2=odd 56 1
int16 Stop bits. 1=1 stop bits, 2=2 stop bits 57 1
uint16 Modbus Address 58 1
uint16 Modbus Port. 0=Disable, 1=Enable 59 1
uint Modbus Timeout (ms) 60 2
char[20] System Time 62 10
char[20] System Date 72 10
int16 Parameter used to request a state machine
change. 1=Not ready, 2=Measure, 3=Zero,
4=Span, 5=Zero Hold, 6=Span Hold, 7=Idle
82 1
char[20] Concentration Unit of Measure for analyte 1 83 10
int16
Float Format index. 0=comma as decimal,
1=period as decimal
93 1
uint16 International date format index. 0=MM\DD\
YYYY, 1=MM-DD-YYYY, 2=DD\MM\YYYY,
3=DD-MM-YYYY
94 1
char[10] Pressure Unit of Measure, case sensitive.
Valid strings include psi, psia, inH20, inHg,
bar, cmH20, kg/cm2, kPa, mbar, micron,
mmHg, oz/in2, torr, atm
95 5
char[10] Temp Unit of Measure, case sensistive. Valid
strings include degreeC, degreeF, degreeK
100 5
uint Data Valid. 0=Not Valid, 1=Valid 105 2
Modbus Registers