4-22 | Model 5100 HD Analyzer
Data Type Description MB Starting
Regs
MB Reg Size
int16 0=Line-lock mode, 1=Sample, 2=Calib 107 1
int16
Pressure and temp compensation. 0=none,
1=pressure, 2=temperature, 3=both
108 1
uint Bit-mapped system alarms 109 2
uint Bit-mapped system alarms 111 2
uint Bit-mapped system alarms 113 2
uint Bit mapped concentration alarms 115 2
int16
System Status. 0=Everything ok, Not 0=some-
thing is wrong
117 1
int16
System State. 100=Not Ready, 201=Measure
Run, 502=Idle, 600=diag
118 1
uint Total uptime in seconds 119 2
uint Uptime in seconds 121 2
oat Cell temp for display 123 2
oat Current Output 1 125 2
oat Current Output 1 O󰀨set 127 2
uint16 Current Output 1 Parameter 129 1
oat Current Output 1 Span 130 2
int16
Current Output 1 Output Type. 0=4-20 mode,
1=0-20 mode.
If overange enable (ANAOUT1OR = 1)
0 = 3.78 - 20.5, 1 = 0-20.5, high or low val-
ues can also be set by fault value, See Fault
option (ANAOUT1F)
132 1
uint16
Current Output 1 Over range Enable. 0=dis-
abled, 1=enabled
133 1
uint16 Current Output 1 Fault Option. 0=No fault,
1=Low fault, 2=High fault.
If high fault (2):
High end of range becomes 22 mA
Output ratio becomes 1
If low fault (1) AND in 4-20 mode (ANAOUT1T
= 0):
Low end of range becomes 3 mA
Output ratio becomes 0.
134 1
oat Current Output 2 - 135 2
oat Current Output 2 O󰀨set - 137 2
uint16
Current Output 2 Parameter-
139 1
oat Current Output 2 Span - 140 2
int16
Current Output 2 Output Type. 0=4-20 mode,
1=0-20 mode.
If overange enable (ANAOUT2OR = 1)
0 = 3.78 - 20.5, 1 = 0-20.5, high or low val-
ues can also be set by fault value, See Fault
option (ANAOUT2F) -
142 1