User Interface | 4-23
Data Type Description MB Starting
Regs
MB Reg Size
uint16
Current Output 2 Over range Enable. 0=dis-
abled, 1=enabled
143 1
uint16
Current Output 2 Fault Option. 0=No fault,
1=Low fault, 2=High fault.
If high fault (2):
High end of range becomes 22 mA
Output ratio becomes 1
If low fault (1) AND in 4-20 mode (ANAOUT2T
= 0):
Low end of range becomes 3 mA
Output ratio becomes 0.
144 1
uint16
Relay 1. 0=OPEN, 1=CLOSED
145 1
uint16
Relay 1 Delay. Holds the relay output for this
set amount of time.
146 1
uint16
Relay 1 Function. 0=Disabled, 1=Concentra-
tion, 3=Data Valid, 4=System alarm
147 1
uint16
Relay 1 Source. The source parameter num-
ber used to evaluated the relay function.
148 1
oat
Relay 1 Value. The value used to evaluate the
relay function.
149 2
uint16
Relay 2. 0=OPEN, 1=CLOSED
151 1
uint16
Relay 2 Delay. Holds the relay output for this
set amount of time.
152 1
uint16
Relay 2 Function. 0=Disabled, 1=Concentra-
tion, 3=Data Valid, 4=System alarm
153 1
uint16
Relay 2 Source. The source parameter num-
ber used to evaluated the relay function.
154 1
oat
Relay 2 Value. The value used to evaluate the
relay function.
155 2
uint16
Relay 3. 0=OPEN, 1=CLOSED
157 1
uint16
Relay 3 Delay. Holds the relay output for this
set amount of time.
158 1
uint16
Relay 3 Function. 0=Disabled, 1=Concentra-
tion, 3=Data Valid, 4=System alarm
159 1
uint16 Relay 3 Source. The source parameter num-
ber used to evaluated the relay function.
160 1
oat Relay 3 Value. The value used to evaluate the
relay function.
161 2
uint16 Relay 4. 0=OPEN, 1=CLOSED 163 1
uint16 Relay 4 Delay. Holds the relay output for this
set amount of time.
164 1
uint16 Relay 4 Function. 0=Disabled, 1=Concentra-
tion, 3=Data Valid, 4=System alarm
165 1
uint16 Relay 4 Source. The source parameter num-
ber used to evaluated the relay function.
166 1
oat