4-24 | Model 5100 HD Analyzer
Data Type Description MB Starting
Regs
MB Reg Size
oat Relay 4 Value. The value used to evaluate the
relay function.
167 2
boolean Master enable for all relays. 0=Disabled,
1=Enabled
169 1
char[20] Analyte 1 Name 170 10
oat Analyte 1 Concentration 180 2
oat Analyte 1 band high limite band 1 182 2
oat Analyte 1 band low limite band 1 184 2
oat Analyte 1 band high limite band 2 185 2
oat Analyte 1 band low limite band 2 188 2
oat Analyte 1 band high limite band 3 190 2
oat Analyte 1 band low limite band 3 192 2
oat Analyte 1 band high limite band 4 194 2
oat Analyte 1 band low limite band 4 196 2
char[20] Analyte 2 Name 198 10
oat Analyte 2 Concentration 208 2
oat TEC Temperature Value 210 2
int16 Heater Actuator Type. 0=Reverse, 1=Forward. 212 1
oat
Heater Deviation Tolerance. Ignored if 0.
Otherwise high or low limit alarm are set when
value is out bounds of 1 deviation
213 2
oat Heater Deriviative Time 215 2
oat Heater Duty Cycle 217 2
oat Heater Integral Time 219 2
oat Heater Proportional Band 221 2
oat Heater Set Point 223 2
oat TEC Sample Time 225 2
oat TEC hysterisi 227 2
oat Electronics Temperature 229 2
oat Cell RTD1 Filtered value 231 2
oat Cell Set Point 233 2
oat Cell Deviation Tolerance. Ignored if 0. Oth-
erwise high or low limit alarm are set when
value is out bounds of 1 deviation
235 2
oat Calibration temperature for concentration
compensation.
237 2
oat Calibration pressure for concentration com-
pensation
239 2
oat Max TEC temperature 241 2
oat Min TEC temperature 243 2
uint Interval in minutes to save the log data 245 2
char[10] Concentration Unit of Measure for analyte 2 247 5