User Interface | 4-25
Data Type Description MB Starting
Regs
MB Reg Size
uint Decimal places to use for displaying analyte 1 252 2
uint Decimal places to use for displaying analyte 2 254 2
boolean Normally closed indicator for relay 1. 1=Nor-
mally closed, 0=Normally open
256 1
boolean Normally closed indicator for relay 2. 1=Nor-
mally closed, 0=Normally open
257 1
boolean Normally closed indicator for relay 3. 1=Nor-
mally closed, 0=Normally open
258 1
boolean Normally closed indicator for relay 4. 1=Nor-
mally closed, 0=Normally open
259 1
uint16 Menu timeout. The time that the menu will
navigate to the homescreen after no activity.
0 is disabled
260 1
uint16 Password timeout. The time that a user
logged into the front display menu will be
logged out after no activity. 0 is disabled
261 1
uint Temporary cache to miror warnings in SYS-
ALARM1
262 2
uint Temporary cache to miror warnings in SYS-
ALARM2
264 2
uint system statuc warning 266 2
uint16 The seed value for all IIR lters. 268 1
oat Max TEC temperature 2 269 2
oat Min TEC temperature 2 271 2
oat Concentration of Reference call 273 2
int16 Pressure (STCELLP) and temp (STCELLT)
static value usage. 0=neither, 1=use static
pressure, 2=use static temperature, 3= use
both
275 1
oat Static cell temperature. Use by setting STAT-
ICPT to 2 or 3
276 2
oat Static cell pressure. Use by setting STAT-
ICPT to 1 or 3
278 2
int16 Analog output 1, 0 - follow source during cal,
1 - hold output during cal
280 1
int16 Analog output 2, 0 - follow source during cal,
1 - hold output during cal L82
281 1
oat Threshold in 2f signal 282 2
oat Reference Laser Power (power at calib) 284 2
oat Laser Sample Period in seconds 286 2
oat Laser Power 288 2
oat Cell length, concentrations are multiplied by
this parameter
290 2
int16 Used to change state to diagnostics from the
menu
292 1