Maintenance | 6-15
System Alarm/Warning bit map
When System Status (parameter 129, 16-bit integer, modbus register 117) contains none-zero value,
the bit map of System Alarms and System Warning (parameter 125, 126, 127, 128; 32-bit integer,
modbus register 109, 110, 111, 112, 113) contains system alarms as defined:
register
type
Parameter ID
TemperatureoftheTEC#1(181)is
outside of a band set by parameter 183
aroundthesetpoint(188)
TemperatureoftheTEC#2(374)is
outside of a band set by parameter 376
aroundthesetpoint(381)
Celltemperature(192)isoutsideofa
band set by parameter 194 around the
Pressuretransducerreading(244)is
outside of a band set by min/max
ForATEXONLY:SEALpressure(366)
is outside of a band set by min/max
File system error, reported in system
alarm(125)
276
TECtemperature (181) drifts out of
factory setting window (197,198)
definedbythefactorydefault(276)
395
TECtemperature2 (381) drifts out of
factory setting window (213,214)
definedbythefactorydefault(395)
Concentration reading from cell #1 is out
of calibration range
Concentration reading from cell #2 is
out of calibration range
Laserintensityinsamplecellislow
Laserintensity in reference cell is low
Concentration in reference cell is low
Concentration alarm from band 1
(channel1)
Concentration alarm from band 2
(channel1)
Concentration alarm from band 3
(channel1)
Concentration alarm from band 4
(channel1)
Concentration alarm from band 1
(channel2)
Concentration alarm from band 2
(channel2)
Concentration alarm from band 3
(channel2)
Concentration alarm from band 4
(channel2)
Concentration reading is invalid.
Possible causes include: system in line-
lock, system in calibration, etc...
System Alarm/Warning bit map
When System Status (parameter 129, 16-bit integer, modbus register 117) con-
tains none-zero value, the bit map of System Alarms and System Warning (pa-
rameter 125, 126, 127, 128; 32-bit integer, modbus register 109, 110, 111, 112,
113) contains system alarms as dened:
Cellpressure(254)isoutsideofaband
setbymin/maxpressure(495,496)
Errorinspancalibration
Errorinzerocalibration
User error detected from the user analog
input
DHCP enable. 1=disabled, 2=enabled
TCPEnable.0=disabled,1=enabled
Version of the params.xml file generated by
this
InternalUART1baudrate(MODBUS)
InternalUART2baudrate(DISPLAY)
ExternalUART1baudrate(RS232)
Interface Board FW version
Parity. 0=none, 1=even, 2=odd
Stop bits. 1=1 stop bits, 2=2 stop bits
ModbusPort.0=Disable,1=Enable
Concentration Unit of Measure for analyte 1
Pressure Unit of Measure, case sensitive.
Valid strings include psi, psia, inH20, inHg,
bar, cmH20, kg/cm2, kPa, mbar, micron,
mmHg, oz/in2, torr, atm
Temp Unit of Measure, case sensistive. Valid
strings include degreeC, degreeF, degreeK
Data Valid. 0=Not Valid, 1=Valid
0=Line-lock mode, 1=Sample, 2=Calib
Bit mapped concentration alarms