Maintenance | 6-15
39
System Alarm/Warning bit map
When System Status (parameter 129, 16-bit integer, modbus register 117) contains none-zero value,
the bit map of System Alarms and System Warning (parameter 125, 126, 127, 128; 32-bit integer,
modbus register 109, 110, 111, 112, 113) contains system alarms as defined:
Modbus
register
Bit
Display message
Alarm
type
Related
Parameter ID
Description
109
1
“TEC Hi Limit Alrm”
Alarm
181, 183, 188
TemperatureoftheTEC#1(181)is
outside of a band set by parameter 183
aroundthesetpoint(188)
109
2
“TEC Lo Limit Alrm”
Alarm
109
3
“TEC2 Hi Limit Alrm”
Alarm
374, 376, 381
TemperatureoftheTEC#2(374)is
outside of a band set by parameter 376
aroundthesetpoint(381)
109
4
“TEC2 Lo Limit Alrm”
Alarm
109
5
“Cell Hi Limit Alrm”
Alarm
192, 193, 194
Celltemperature(192)isoutsideofa
band set by parameter 194 around the
setpoint(193)
109
6
“Cell Lo Limit Alrm”
Alarm
109
7
“Pres Hi Limit Alrm
Alarm
244, 236, 237
Pressuretransducerreading(244)is
outside of a band set by min/max
pressure(236,237)
109
8
“Pres Lo Limit Alrm”
Alarm
109
9
“Seal Hi Limit Alrm”
Alarm
366, 369, 370
ForATEXONLY:SEALpressure(366)
is outside of a band set by min/max
pressure(369,370)
109
10
“Seal Lo Limit Alrm”
Alarm
109
11
"File error"
Alarm
File system error, reported in system
alarm(125)
109
12
"Internal error"
Alarm
Internal harware error
111
13
“TEC temp drift”
Warning
181, 197, 198,
276
TECtemperature (181) drifts out of
factory setting window (197,198)
definedbythefactorydefault(276)
111
14
“TEC2 temp drift”
Warning
381, 213, 214,
395
TECtemperature2 (381) drifts out of
factory setting window (213,214)
definedbythefactorydefault(395)
111
15
"Conc.Outofrange"
Warning
Concentration reading from cell #1 is out
of calibration range
111
16
"Conc.Outofrange2"
Warning
Concentration reading from cell #2 is
out of calibration range
111
17
"LowlightinSample"
Warning
Laserintensityinsamplecellislow
111
18
"Battery failed"
Warning
Internal battery failure
111
19
"LowlightinRef"
Warning
Laserintensity in reference cell is low
111
20
"Refconcalarm"
Warning
Concentration in reference cell is low
111
21
"Conc. Band 1"
Warning
Concentration alarm from band 1
(channel1)
111
22
"Conc. Band 2"
Warning
Concentration alarm from band 2
(channel1)
111
23
"Conc. Band 3"
Warning
Concentration alarm from band 3
(channel1)
111
24
"Conc. Band 4"
Warning
Concentration alarm from band 4
(channel1)
111
25
"Conc. Band 1"
Warning
Concentration alarm from band 1
(channel2)
111
26
"Conc. Band 2"
Warning
Concentration alarm from band 2
(channel2)
111
27
"Conc. Band 3"
Warning
Concentration alarm from band 3
(channel2)
111
28
"Conc. Band 4"
Warning
Concentration alarm from band 4
(channel2)
111
29
"Data Invalid"
Warning
122, 363
Concentration reading is invalid.
Possible causes include: system in line-
lock, system in calibration, etc...
System Alarm/Warning bit map
When System Status (parameter 129, 16-bit integer, modbus register 117) con-
tains none-zero value, the bit map of System Alarms and System Warning (pa-
rameter 125, 126, 127, 128; 32-bit integer, modbus register 109, 110, 111, 112,
113) contains system alarms as dened:
40
111
30
“Cell pres Hi”
Warning
254, 495, 496
Cellpressure(254)isoutsideofaband
setbymin/maxpressure(495,496)
111
31
“Cell pres Low”
Warning
111
32
"Span error"
Warning
Errorinspancalibration
112
1
"Zero error"
Warning
Errorinzerocalibration
113
2
"User error"
Warning
User error detected from the user analog
input
Common Modbus Registers
ID
Access
Data
Type
mnemonic
MB
Starting
Regs
MB Reg
Size
Description
6
rs
char[20]
SERIAL
0
10
Serial Number
7
rs
char[6]
MAC
10
3
MACaddress
8
rws
uint
IP
13
2
IPAddress
9
rws
uint
SUBNET
15
2
IP Subnet Mask
10
rws
uint
GW
17
2
Gateway
11
rws
uint
DNS
19
2
DNS
12
rws
uint16
DHCPE
21
1
DHCP enable. 1=disabled, 2=enabled
13
rws
uint16
TCPE
22
1
TCPEnable.0=disabled,1=enabled
14
rws
uint
SVRIP
23
2
Server IP
100
rwf
int16
CFGFILEVER
25
1
Version of the params.xml file generated by
this
101
r
char[20]
VERSION
26
10
Software Version
102
r
char[20]
NAME
36
10
Model Name
103
rw
uint
IU1BAUD
46
2
InternalUART1baudrate(MODBUS)
104
rwf
uint
IU2BAUD
48
2
InternalUART2baudrate(DISPLAY)
105
rwf
uint
EU1BAUD
50
2
ExternalUART1baudrate(RS232)
106
rwf
uint
EU2BAUD
52
2
ExternalUART2baudrate
107
r
uint16
FWVERHRB
54
1
Interface Board FW version
108
r
uint16
CARDID
55
1
0x20: HC121, 0x40: HC122
109
rw
int16
MBPARITY
56
1
Parity. 0=none, 1=even, 2=odd
110
rwf
int16
MBSTOPBIT
57
1
Stop bits. 1=1 stop bits, 2=2 stop bits
111
rw
uint16
MBUSADD
58
1
ModbusAddress
112
rwf
uint16
MBUSPORT
59
1
ModbusPort.0=Disable,1=Enable
113
rwf
uint
MBUSTO
60
2
ModbusTimeout(ms)
114
r
char[20]
SYSTIME
62
10
System Time
115
r
char[20]
SYSDATE
72
10
System Date
117
rw
char[20]
CONCUNIT
83
10
Concentration Unit of Measure for analyte 1
120
rw
char[10]
PRESUNIT
95
5
Pressure Unit of Measure, case sensitive.
Valid strings include psi, psia, inH20, inHg,
bar, cmH20, kg/cm2, kPa, mbar, micron,
mmHg, oz/in2, torr, atm
121
rw
char[10]
TEMPUNIT
100
5
Temp Unit of Measure, case sensistive. Valid
strings include degreeC, degreeF, degreeK
122
r
uint
DATAVALID
105
2
Data Valid. 0=Not Valid, 1=Valid
123
r
int16
DMODE
107
1
0=Line-lock mode, 1=Sample, 2=Calib
125
r
uint
SYSALARM1
109
2
Bit-mapped system alarms
126
r
uint
SYSALARM2
111
2
Bit-mapped system alarms
127
r
uint
USERALARM
113
2
Bit-mapped system alarms
128
r
uint
CONCALARM
115
2
Bit mapped concentration alarms