6-16 | Model 5100 HD Analyzer
Cellpressure(254)isoutsideofaband
setbymin/maxpressure(495,496)
Errorinspancalibration
Errorinzerocalibration
User error detected from the user analog
input
Type
Starting
Size
DHCP enable. 1=disabled, 2=enabled
TCPEnable.0=disabled,1=enabled
Version of the params.xml file generated by
this
InternalUART1baudrate(MODBUS)
InternalUART2baudrate(DISPLAY)
ExternalUART1baudrate(RS232)
Interface Board FW version
Parity. 0=none, 1=even, 2=odd
Stop bits. 1=1 stop bits, 2=2 stop bits
ModbusPort.0=Disable,1=Enable
Concentration Unit of Measure for analyte 1
Pressure Unit of Measure, case sensitive.
Valid strings include psi, psia, inH20, inHg,
bar, cmH20, kg/cm2, kPa, mbar, micron,
Temp Unit of Measure, case sensistive. Valid
strings include degreeC, degreeF, degreeK
Data Valid. 0=Not Valid, 1=Valid
0=Line-lock mode, 1=Sample, 2=Calib
Bit mapped concentration alarms
Common Modbus Registers
System Status. 0=Everythingok,Not
0=something is wrong
SystemState.100=NotReady,201=Measure
Run,502=Idle,600=diag
CurrentOutput1Parameter
CurrentOutput1OutputType.0=4-20 mode,
1=0-20 mode.
Ifoverangeenable(ANAOUT1OR=1)
0 = 3.78 - 20.5, 1 = 0-20.5, high or low
values can also be set by fault value, See
Faultoption(ANAOUT1F)
CurrentOutput1OverrangeEnable.
0=disabled, 1=enabled
CurrentOutput1FaultOption.0=Nofault,
1=Lowfault,2=Highfault.
Ifhighfault(2):
Highendofrangebecomes22mA
Outputratiobecomes1
Iflowfault(1)ANDin4-20 mode
(ANAOUT1T=0):
Lowendofrangebecomes3mA
Outputratiobecomes 0.
CurrentOutput2Offset-
CurrentOutput2Parameter-
CurrentOutput2OutputType.0=4-20 mode,
1=0-20 mode.
Ifoverangeenable(ANAOUT2OR=1)
0 = 3.78 - 20.5, 1 = 0-20.5, high or low
values can also be set by fault value, See
Faultoption(ANAOUT2F)-
CurrentOutput2OverrangeEnable.
0=disabled, 1=enabled
CurrentOutput2FaultOption.0=Nofault,
1=Lowfault,2=Highfault.
Ifhighfault(2):
Highendofrangebecomes22mA
Outputratiobecomes1
Iflowfault(1)ANDin4-20 mode
(ANAOUT2T=0):
Lowendofrangebecomes3mA
Outputratiobecomes0.
Relay1.0=OPEN,1=CLOSED
Relay1Delay.Holdstherelayoutputforthis
set amount of time.
Relay1Function.0=Disabled,
1=Concentration, 3=Data Valid, 4=System
alarm
Relay1Source.Thesourceparameter
number used to evaluated the relay function.
Relay1Value.Thevalueusedtoevaluate
the relay function.
Relay2.0=OPEN,1=CLOSED