2-2 | Model 5100 HD Analyzer
Parameter Specication
Wavelength
Depends on analyte and availability:
H2O: 1854nm (25789JE)
CO2: 2004 nm (25834JE)
O2: 763 nm (25895JE)
H2S: 1575 nm (25881JE)
CO2 (high range): 1581 nm (25889JE)
Laser Classication
25789JE, Class 1m
25834JE, Class 1m
763 nm, Class 3b
1575 nm, Class 3b
1581 nm, Class 3b
Dynamic Range
324 mAU (milli-Absorbance Units) for 0 - 2500
ppmV H2O
1.015 AU for 0 - 10% CO2
268 mAU for 0 - 20.95 % O2
18 mAU for 0 - 5% H2S
Limit Of Detection (LOD)
H2O in CH4: 4 ppmV (0.53 mAU)
H2O in non-absorbing background gas (at 1854
nm): 1 ppmV (.16 mAU)
CO2 in non-absorbing background gas (at 2004
nm): 100 ppmV (1.0 mAU)
O2 in non-absorbing background gas (at 760 nm):
250 ppmV (.33 mAU)
H2S in non-absorbing background gas (at 1575 nm):
1000 ppmV (.36 mAU)
Number of analytes 1 or 2, depending on conguration
Number of sample points 1
Speed of response < 2 seconds (photometric); < 50 seconds to T90
including sample system purge (At ow rate of 2
liters/minute)
Accuracy H2O in CH4: +/- 4 ppmV (.53 mAU) or +/- 2%,
whichever is greater
H2O in non-absorbing background: +/- 1 ppmV (.16
mAU) or 2% of reading, whichever is greater
CO2 in non-absorbing background: +/- .1% CO2 (10
mAU) or 2% of reading, whichever is greater
O2 in non-absorbing background: +/- .1% O2 (1.32
mAU) or 2% of reading, whichever is greater
H2S in non-absorbing background: +/- .1% (.36
mAU) or 2% of reading,
whichever is greater