Caution…
About the use:
Do not use the instrument if the ventilator is out of
order.
Before cleaning the calibrator, you must switch it off,
allow it to cool down and remove all cables.
About the well, insertion tube and grid plate:
The well and the insertion tube must be clean before
use.
Scratches and other damage to the insertion tubes
should be avoided by storing the insertion tubes
carefully when not in use.
The insertion tube must never be forced into the
well. The well could be damaged as a result, and the
insertion tube may get stuck.
Before using new insertion tubes for the calibration,
the insertion tubes must be heated up to maximum
temperature - 320°C (608°F) / 650°C (1202°F) - for a
period of minimum 30 minutes.
In order to ensure the best calibration of your
sensors please avoid using insertion tubes in the
ITC-320 calibrator which have been used in the ITC-
650 calibrator.
The insertion tube must always be removed from the
calibrator after use.
The humidity in the air may cause verdigris to form
on the insertion tube inside the instrument. There is
a risk that the insertion tube may become stuck if
this is allowed to happen.
If the calibrator is to be transported, the insertion
tube must be removed to avoid damage to the
instrument.

Other manuals for Ametek JOFRA ITC-320 A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek JOFRA ITC-320 A and is the answer not in the manual?

Ametek JOFRA ITC-320 A Specifications

General IconGeneral
BrandAmetek
ModelJOFRA ITC-320 A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals