4-LEAD ELECTROCHEMICAL SENSOR EXAMPLE
Many electrochemical sensors come in 4-lead packages that have a
counter, a reference, and two sensing electrodes. The ADuCM355
supports biasing and measuring of these sensor types.
The M355_ECSns_DualWE example project configures the
low power, potentiostat CH0 channel to bias the sensor. The
current flowing to and from the SE0 pin is measured via the
low power TIA Channel 0 (TIA0). The current flowing from
the SE1 electrode is measured via the low power TIA Channel 1
(TIA1).
The TIA amplifiers convert the current to a voltage that is
measured via the analog-to-digital converter (ADC), and the
source code calculates the current flowing in each electrode.
The M355_ECSns_DualWE code example project is located in
the examples folder.
Figure 30 shows the configurable parameters located in the
AD5940Main.c file. Modify the value of the correct
LpTiaRtiaSel parameter for each channel based on the
maximum expected current.
Figure 31 shows the connection details between the 4-lead
sensor and the ADuCM355.