What to do if Applied Biosystems Laboratory Equipment has no beep?
- TThomas ClaySep 6, 2025
For the Applied Biosystems Laboratory Equipment, check Run-Time Beep configuration.
What to do if Applied Biosystems Laboratory Equipment has no beep?
For the Applied Biosystems Laboratory Equipment, check Run-Time Beep configuration.
Why does my Applied Biosystems GeneAmp PCR System 9700 printer fail?
If your Applied Biosystems Laboratory Equipment printer fails, verify the printer settings (baud rate = 9600, no parity, one stop bit, eight data bits) and ensure you are using the correct Applied Biosystems printer cable.
Why is my Applied Biosystems GeneAmp PCR System 9700 making too much noise?
If the Applied Biosystems Laboratory Equipment is making too much noise, check for any obstructions in the side vents.
What to do if Applied Biosystems Control Unit displays 'Remove infinity hold first'?
To resolve this issue on Applied Biosystems Control Unit, add segments prior to the post-PCR infinity hold.
What to do if Applied Biosystems GeneAmp PCR System 9700 cooling rate is too slow?
Move the Applied Biosystems Laboratory Equipment to a well-ventilated location (15-30°C).
What to do if Applied Biosystems Control Unit has no screen display and no response when turned on?
To resolve this issue with Applied Biosystems Control Unit, ensure the power switch is ON, the power cord is connected, and check the fuses.
What to do if Applied Biosystems GeneAmp PCR System 9700 displays 'Method requires at least one segment'?
To resolve the 'Method requires at least one segment' error on your Applied Biosystems Laboratory Equipment, review and correct the method to include the temperature parameter(s).
How to fix 'No seconds in time field' error on Applied Biosystems GeneAmp PCR System 9700?
To resolve the 'No seconds in time field' error on your Applied Biosystems Laboratory Equipment, include seconds when entering the time.
What to do if Applied Biosystems Control Unit displays message 'Valid range is'?
To solve this problem with Applied Biosystems Control Unit, reenter a value within the parameters of the field.
Why Applied Biosystems GeneAmp PCR System 9700 cooling fan does not make whirring sound?
To fix the issue where the Applied Biosystems Laboratory Equipment instrument cooling fan does not make a whirring sound, verify that the power switch is on, the power cord is connected, and check the fuses.
Temperature Range | 4°C to 99.9°C |
---|---|
Temperature Accuracy | ±0.25°C |
Temperature Uniformity | ±0.5°C |
Block Format | 96-well |
Temperature Uniformity at 60°C | ±0.5°C |
Display | LCD |
Power Requirements | 100-240V, 50/60Hz |
Sample Capacity | 96 samples |
Sample Volume Range | 10-100 µL |