Code Description
Process to Clear
Press EXIT key Wait 5sec,
Press EXIT key
Reboot. If error
message repeats
contact manufacturer
Contact
Manufacturer
VSENSE_INT_OVP
Displayed when the output
voltage measurement is
greater than the output
voltage setting by 5V in the
Lo range and 10V in the
Hi range. Displayed when
the System Parameter Volt
Sense=INT
•
VSENSE_INT_LVP
Displayed when the output
voltage measurement is
greater than the output
voltage setting by 5V in the
Lo range and by 10V in the
Hi range. Displayed when
the System Parameter Volt
Sense=EXT
•
VSENSE_EXT_OVP
Displayed if the output current
exceeds the peak (Inrush)
output current by 10% in the
Lo range (0-155V) and by
20% in the Hi range (0-310V).
•
VSENSE_EXT_LVP
Displayed when the output
voltage measurement is
less than the output voltage
setting by 5V in the Lo range
and by 10V in the Hi range.
Displayed when the System
Parameter Volt Sense=EXT
•
FAN_FAIL
Displayed when the cooling
fan(s) fail.
• •
DA_OTP
Displayed when the
temperature of the AMP
board reaches 100°C.
• •
DD_FAIL
Displayed when the DD board
internal check fails.
• •
INT_TEST_FAIL
Displayed when the voltage
measurement is different from
the setting by ±20V during
internal verification.
•
A-Hi
Displayed when the output
current exceeds the A-Hi test
parameter setting.
•
A-Lo
Displayed when the output
current stays below the A-Lo
test parameter setting.
(List Mode only)
•
OUTPUT_SHORT
Displayed when a short is
detected on the outputs.
•