61
BASE KEYWORD DESCRIPTION VALUE
MANual :FILE:ADD <file_name> Add File
0-9, A – Z, Limit 23 characters
for filename.
:FILE:EDIT|OPEN <file_name>
:EDIT|OPEN?
Load a File
:FILE:COPY <file_source>,<file_destination> Copy File
:FILE:DELete <file_name> Delete File 0.0 – 420.0 VDC
:FILE:TOTal? Query File Total
:FILE:INDex <value>
:FILE:INDex?
Set File Index
:FILE:NAME? Query File Name
:COUPle <value>
:COUPle?
Set Coupling Mode AC, DC, ACDC
:WAVE
:WAVE?
Set Output Waveform
SINE, TRIangle, SQUare,
CLIPped
:THD <value>
:THD?
Set THD Value *Clipped wave only
:RANGe
:RANGe?
Set Voltage Range AUTO, HIGH, LOW
:VOLTage:AC <value>
:VOLTage:AC?
Set Output AC Voltage 0.0 – 310.0 VAC
:VOLTage:DC <value>
:VOLTage:DC?
Set Output DC Voltage 0.0 – 420.0 VDC
:FREQuency
:FREQuency?
Set Frequency 5 – 1200 Hz
:RAMP:UP <value>
:RAMP:UP?
Set Ramp Up 0.1 – 999.9s, 0 = OFF
:CURRent[:LIMit]:HIGH <value>
:CURRent[:LIMit]:HIGH?
Set Current High Limit *See Specifications
:CURRent[:LIMit]:DELay <value>
:CURRent[:LIMit]:DELay?
Set Delay *See Specifications
POWer[:LIMit]:HIGH <value>
:POWer[:LIMit]:HIGH?
Set Power High Limit *See Specifications
ANGLe[:STARt] <value>
:ANGLe[:STARt]?
Set Start Angle 0 – 359°
A) Manual Mode Commands and Companion Queries
These commands are used to modify test parameters for the Manual mode on the instrument. The
operating mode of the unit should be put in Manual mode for the commands below to work successfully.
These commands require a parameter value to be included with the command. The companion query
command will read the parameter using the same value that is used for setting the parameter.
7.4.3 Mode and File Subsystem