© Ikonix USA
68
BASE KEYWORD DESCRIPTION VALUE
:CREStfactor? Query Crest Factor 0.0 – 10.00
:APParent? Query Apparent Power *See Specifications
System Commands and Companion Queries
These commands are used to modify the system parameters for the instrument. The parameters below
control the global system parameter that applies in all Output Modes of this power supply. These
commands require a parameter value to be included with the command. The companion query command
will read the parameter using the same value that is used for setting the parameter.
7.4.6 System Subsystem
BASE KEYWORD DESCRIPTION VALUE
SYSTem
:PLC[:REMote] <value>
Query Crest Factor
Set PLC 0, 1, ON, OFF
:POWUP <value>
:POWUP?
Set Power Up State OFF, ON, LAST
:ALARm <value>
:ALARm?
Alarm Volume 0 – 9 , 0 = OFF
:OCFold <value>
:OCFold?
Set OCFold Back 0, 1, ON, OFF
:SENSe[:VOLTage] <value>
:SENSe[:VOLTage]?
Set Voltage Sense INTernal, EXTernal
:SYNChronous[:SIGNal] <value>
:SYNChronous[:SIGNal]?
Set SYNC Signal
START, END, BOTH, OFF, ON,
EVENT
[:LIMit]:VOLTage[:AC]:HIGH <value>
[:LIMit]:VOLTage[:AC]:HIGH?
Set System AC Voltage High
Limit
0.0 – 310.0 VAC
[:LIMit]:VOLTage[:AC]:LOW <value>
[:LIMit]:VOLTage[:AC]:LOW?
Set System AC Voltage Low
Limit
0.0 – 310.0 VAC
[:LIMit]:VOLTage:DC:HIGH <value>
[:LIMit]:VOLTage:DC:HIGH?
Set System DC Voltage High
Limit
0.0 – 420.0 VDC
[:LIMit]:VOLTage:DC:LOW <value>
[:LIMit]:VOLTage:DC:LOW?
Set System DC Voltage Low
Limit
0.0 – 420.0 VDC
[:LIMit]:VOLTage:FREQuency:HIGH <value>
[:LIMit]:VOLTage:FREQuency:HIGH?
Set System Frequency High Limit 5.0 – 1200 Hz
[:LIMit]:VOLTage:FREQuency:LOW <value>
[:LIMit]:VOLTage:FREQuency:LOW?
Set System Frequency Low Limit 5.0 – 1200 Hz
:LOCK <value>
:LOCK?
Set Lock 0, 1, ON, OFF
:GPIB <value>
:GPIB?
Set GPIB Address
:LAN:ADDRess
:LAN:ADDRess?
Set IP Address
:MAC? Query MAC ID