AQUA SALT
+
MAINTENANCE AND INSTRUCTION MANUAL
Salt chlorine genera ENGLISH
ADSP7000753 rev. 5.0 09/07/2019 23/51
6) Wait without touching the probe or its cable for at least 60 seconds, until the value read is
stabilized, then press Enter.
7) At the end of the countdown or after pressing Enter, the pH value of the solution (4.01) and the
value read by the probe (in mV) will be shown.
8) The procedure ends visualizing the quality of the probe with a percentage from 0 to 100. If
quality is below 25%, we recommend to replace the probe.
9) Place the probe back in the probe holder or on the piping of the filtration system.
• Press Enter to move through the calibration steps.
• Press and hold Enter for about 1 second to go back to the pH Calibration section of the Installer
Menu.
The Programming menu sets the parameters to programme and manage the operation of your chlorine
generator.
• Press F to scroll the menu sections
• Press and hold Enter for 1 second to go back to the Programming section of the Installer Menu.