For best results, AZI recommends that the instrument is provided
with its own DEDICATED electrical power outlet. Do not put it on a
circuit with motors, blenders, heaters, coolers, grinders, or other high
current electrical devices.
The power source outlet that is used by the MAX
®
4000 MUST BE
GROUNDED. (If you are not sure if the socket is grounded, check
with your plant electrician before proceeding). The MAX
®
4000 will
appear to function correctly even if the socket is not grounded,
but there is a danger of possible electrocution! If grounded
outlets are not available, consider using a ground fault
interrupter to protect personnel against electrical shock.
• Plug the power cord into a dedicated electrical power outlet. Line noise on the power
lines may affect accuracy. A dedicated AC power line or an isolation transformer for
the MAX
®
4000 may be beneficial in eliminating this noise.
• Turn on the instrument by flipping the switch on the power entry module on the back
of the instrument; the instrument is designed to remain powered up at all times.
• From a cold start, allow the instrument to warm-up for 60 minutes before testing.