TROUBLESHOOTING AND ALARMS
49
Continue on the next page.
INDICATOR POSSIBLE CAUSE SOLUTION ALARM
ACTIVATION
DELAY
POWER
FAILURE
Power source failure. Check if power is available and
check the power cable.
When the battery is fully
charged the Auralis pump works
for a minimum of 3 hours.
Less than 2 seconds
BATTERY LOW
Less than 1 hour left on the
battery.
Charge the battery by using
mains power supply to run the
Auralis pump.
500 ms
READY FOR
SERVICE
and
Front panel
alarms show
green
The Auralis pump needs
service.
&RQWDFWTXDOL¿HGVHUYLFH
personnel, quoting the serial
number.
7R¿QGWKHVHULDOQXPEHUVIRU
the Auralis pump, mattress
or seat cushion see sections
³/DEHOV´RQSDJHDQG
³/DEHOVRQ$XUDOLVPDWWUHVVHV
Auralis seat cushion and Auralis
PDWWUHVVH[WHQVLRQV´RQSDJH
58
.
Immediate upon
expiry of service timer
INTERNAL
FAULT
and
Front panel
alarms show
yellow
The Auralis pump has
detected an internal fault.
Switch the Auralis pump off
and restart it. If alarm persists,
FRQWDFWTXDOL¿HGVHUYLFH
personnel.
Maximum 22 minutes
AUTOFIRM
does not
activate
1. The Auralis seat cushion
is in use.
2. The Auralis mattress is in
use.
1. Disconnect the Auralis seat
cushion. 7KH$XWR¿UPPRGH
only applies to the Auralis
mattresses.
2. 7RDFWLYDWH$XWR¿UPPRGH
SUHVVDQGKROGWKH$XWR¿UP
button for 2 seconds.
Not applicable