TECHNICAL SPECIFICATIONS
53
TRANSPORT AND STORAGE CONDITIONS
Short term (Up to 30 days):
Temperature (Ambient) -20 °C to 60 °C (-4 °F to 140 °F)
Relative humidity range 0 % to 95 %
Long term (> 30 days):
Temperature (Ambient) 0 °C to 40 °C (32 °F to 104 °F)
Relative humidity range 0 % to 95 % (non-condensing)
CAUTION
To avoid damage to the Auralis system:
• Do not store the Auralis system in direct sunlight.
• Store the
Auralis pump and Auralis mattress in the protective bags supplied.
• Do not store the
Auralis pump at high temperatures for extended periods. This may cause potential
battery damage.
• Do not store the Auralis pump longer than 5 months without recharging the battery.
AURALIS MATTRESSES AND SEAT CUSHION MEASUREMENTS
Part no Description Size mm (in) Weight kg (Ib)
(Including transport bag of weight 0.46kg)
636T01S Standard 110 2030 x 860 x 115 (80 x 34 x 4.5) 7.9 (17.4)
636T01N Narrow 110 2030 x 780 x 115 (80 x 30 x 4.5) 7.7 (16.9)
636M05S Standard 175 2030 x 860 x 175 (80 x 34 x 7) 13.8 (30.4)
636M05N Narrow 175 2030 x 780 x 175 (80 x 30 x 7) 12.7 (27.9)
636M02S Standard 200 2030 x 860 x 205 (80 x 34 x 8) 11.3 (24.9)
636M02N Narrow 200 2030 x 780 x 205 (80 x 30 x 8) 11 (24.2)
636C01S Seat cushion 470 x 455 x 50 (18.5 x 17.9 x 2) 1.2 (2.6)
636B02 Plus 2140 x 1220 x 205 (84 x 48 x 8) 16 (35.3)
Cover material Premium fabric
ALLOWED COMBINATIONS
Auralis • Skin IQ
®
MCM, Coverlet Only
• Skin IQ
®
365, Coverlet Only
Auralis Plus • Skin IQ
®
1000, Coverlet Only
EXPECTED SERVICE LIFE
Auralis pump 7 years
SOUND LEVELS
Sound level during audible alarm 55 dBA at 1.8 m (1 m horizontal, 1.5 m vertical)