Operating the Maxi Sky 2 Plus in Dual Mode 15
Movement Controls
Operating the Maxi Sky 2 Plus is achieved with the
aid of the command buttons on the hand control.
Vertical Movement
The and buttons are used to raise or
lower the patient. When either button is
pressed, both straps on the Maxi Sky 2 Plus
will move together.
NOTE: Although the ascending and descending
speeds are synchronized by software, a number
of factors may cause the two spreader bars to
move at slightly different speeds, which in turn may
cause minor changes in the patient's nal position.
Readjust as necessary.
The raising and lowering functions
accelerate gradually to prevent any abrupt
movements when the button is initially
pressed, and remain active for as long as
the button remains pressed.
The movement decelerates gradually to
prevent any abrupt movement when the
button is released.
NOTE: If the button is continuously pressed,
the ceiling lifts in the Maxi Sky 2 Plus will each
reach their highest raising limits, but at different
times. This will produce a tilting motion that will
cause the patient's position to shift.
Horizontal Movement and Rotation
Horizontal movement, as well as rotating the patient,
with the Maxi Sky 2 Plus is achieved by manually
pulling or pushing the lift to the desired location (see
Fig. 17).
Maxi Sky
2 Plus Patient Tilt Function
The Maxi Sky 2 Plus is equipped with a patient tilt
function. The hand control features the
and
buttons, which allows for the positioning of the
patient in the upright or reclined position according
to the transfer starting and/or destination position.
Press and hold to tilt patient to a reclined
position.
Press and hold to bring patient up to sitting
position.
The Maxi Sky 2 Plus will activate only the leading
ceiling lift when adjusting the patient’s position.
Therefore, no matter which one of the patient
positioning buttons are pressed, only the “foot-end”
spreader bar will rise or lower accordingly. The
travel of this spreader bar is limited by software to
350 mm above and below the "head-end" spreader
bar.