001-31249-EN rev. 3
TechnicalSpecications48
Electromagnetic Emissions
guidance and manufacturers declaration
electromagnetic emissions - for all eQuipment and systems
The Maxi Sky 2 Plus is intended for use in the electromagnetic environment indicated below. The customer or the user of
the
Maxi Sky 2 Plus should ensure that it is used in such an environment.
Emissions test Compliance
Electromagnetic environment -
guidance
RF emissions CISPR 11 Group 1
The
Maxi Sky
2 Plus uses RF energy
only for its internal function. Therefore,
its RF emissions are very low and are
not likely to cause any interference in
nearby electronic equipment.
RF emissions CISPR 11 Class B The
Maxi Sky
2 Plus is suitable for use
in all establishments, including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage uctuations/ icker emissions
IEC 61000-3-3
Complies
Electromagnetic Immunity
Electromagnetic Immunity
guidance and manufacturers declaration
electromagnetic immunity - for all eQuipment and systems
The
Maxi Sky
2 Plus is intended for use in electromagnetic environment specied below. The customer or the user of the
Maxi Sky
2 Plus should ensure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level
Electromagnetic
environment-guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be wood,
concrete or ceramic tile. If oors
are covered with synthetic
material, the relative humidity
should be at least 30%.
Electrical fast transient/burst
IEC 61000-4-4
±2 kV, AC Mains
±1 kV, I/O Ports
100 kHz repetition frequency
±2 kV, AC Mains
±1 kV, I/O Ports
100 kHz repetition frequency
Mains power quality should
be that of a typical
commercial or hospital
environment.
Surge
IEC 61000-4-5
±2 kV, AC Mains, Line to
Ground
±1 kV, AC Mains, Line to Line
±2 kV, AC Mains, Line to
Ground
±1 kV, AC Mains, Line to Line
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
0 % U
T
; 0,5 cycle
At 0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°
0 % U
T
; 1 cycle
and
70 % U
T
; 25/30 cycles
Single phase: at 0°
0 % U
T
; 250/300 cycle
0 % U
T
; 0,5 cycle
At 0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°
0 % U
T
; 1 cycle
and
70 % U
T
; 25/30 cycles
Single phase: at 0°
0 % U
T
; 250/300 cycle
Power frequency
(50/60 Hz)
magnetic eld
IEC 61000-4-8
30 A/m
50/60 Hz
30 A/m
50/60 Hz
Power frequency magnetic
elds should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
NOTE: U
T
is the AC mains voltage prior to application of the test level.