15A3356O
APPENDIX B (continued)
Sample Pipette Calibration Printout Sample Pipette Calibration Printout
(Artel Nomenclature) (ISO Nomenclature)
(Optional Uncertainty Included) (Optional Uncertainty Omitted)
(Optional Signature Block Included) (Optional Signature Block Omitted)
(Your Lab's Name)
Date: 01 Jun 2011
Time: 15:18
PIPETTE CALIBRATION
Pipette ID: 67443
Operator ID: 39
Reagent Lot Code: 18356
Pipette Sample Volume: 20.00 uL
Sample Solution Range: 3
SAMPLE # VOLUME
1 20.06
2 20.00
3 20.05
4 20.10
5 20.08
6 20.10
7 20.07
8 20.12
9 20.16
10 20.09
Number of Data Points: 10
Mean Volume: 20.08 uL
Absolute Inaccuracy: 0.08 uL
Relative Inaccuracy: 0.41 %
Standard Deviation: 0.043 uL
Coef. of Variation (CV): 0.21 %
Uncertainty of Mean,k=2: 0.1200 uL
Temperature: 21.2 C
Last Instr. Cal. Date: 01 Jun 2011
Last Instr. Cal. No.: 105
Instrument Serial No.: 8282
Software Version: PCS 7A2.203
________________________ ____________
Performed by Date
________________________ ____________
Reviewed by Date
ARTEL PCS PIPETTE CALIBRATION SYSTEM
Patents:
USA: 5492673, 5298978, 7772008, 7791716
Europe: EP0711404, EP0628157
Japan: 2520852
(Your Lab's Name)
Date: 01 Jun 2011
Time: 15:18
PIPETTE CALIBRATION
Pipette ID: 67443
Operator ID: 39
Reagent Lot Code: 18356
Pipette Sample Volume: 20.00 uL
Sample Solution Range: 3
SAMPLE # VOLUME
1 20.06
2 20.00
3 20.05
4 20.10
5 20.08
6 20.10
7 20.07
8 20.12
9 20.16
10 20.09
Number of Data Points: 10
Mean Volume: 20.08 uL
Systematic Error: 0.08 uL
% Systematic Error: 0.41 %
Random Error: 0.043 uL
% Random Error: 0.21 %
Temperature: 21.2 C
Last Instr. Cal. Date: 01 Jun 2011
Last Instr. Cal. No.: 105
Instrument Serial No.: 8282
Software Version: PCS 7A2.203
ARTEL PCS PIPETTE CALIBRATION SYSTEM
Patents:
USA: 5492673, 5298978, 7772008, 7791716
Europe: EP0711404, EP0628157
Japan: 2520852