Loading...
15A3356O
APPENDIX B (continued)
Sample Pipette Calibration Printout for
Multiple Volumes on Variable Volume Pipette
Header
Results for first volume
Statistical Summary for first volume
Results and Statistical Summary
Repeated for Each Additional
Volume
Statistical Summary
Footer
(Your Lab's Name)
Date: 01 Jun 2011
Time: 15:25
PIPETTE CALIBRATION
Pipette ID: 67443
Operator ID: 39
Reagent Lot Code: 18356
Pipette Sample Volume: 20.00 uL
Sample Solution Range: 3
SAMPLE # VOLUME
1 20.06
2 20.00
3 20.05
4 20.10
5 20.08
6 20.10
7 20.07
8 20.12
9 20.16
10 20.09
Number of Data Points: 10
Mean Volume: 20.08 uL
Absolute Inaccuracy: 0.08 uL
Relative Inaccuracy: 0.41 %
Standard Deviation: 0.043 uL
Coef. of Variation (CV): 0.21 %
Temperature: 21.2 C
Pipette Sample Volume: 50.00 uL
Sample Solution Range: 2
SAMPLE # VOLUME
1 50.16
2 50.07
3 50.12
4 50.21
5 50.08
6 50.12
7 50.26
8 50.18
9 50.09
10 50.22
Number of Data Points: 10
Mean Volume: 50.15 uL
Absolute Inaccuracy: 0.15 uL
Relative Inaccuracy: 0.30 %
Standard Deviation: 0.065 uL
Coef. of Variation (CV): 0.13 %
Temperature: 21.3 C
Last Instr. Cal. Date: 01 Jun 2011
Last Instr. Cal. No.: 105
Instrument Serial No.: 8282
Software Version: PCS 7A2.203
ARTEL PCS PIPETTE CALIBRATION SYSTEM
Patents:
USA: 5492673, 5298978, 7772008, 7791716
Europe: EP0711404, EP0628157
Japan: 2520852

Table of Contents