③ Transmit data
Every time a measurement is taken, a new row of data appears in the Tera Term window.
Fixed range (No zero suppression) : ○○.○○,△△△.△ CR / LF
○○.○○ : DMF(%)
The DMF(%) display range is "-2.00 to 40.50".
Example
Display DMF(%)
06.30 6.30
25.00 25.0
LL.LL -2.01 or less
HH.HH 40.51 or higher
EE.EE Error
△△△.△ : Temperature
The temperature output range is "-15.5 to 160.5" in the case of ℃, and is
"004.1 to 320.9" in the case of °F.
Example
℃ °F
Output Data Temperature Output Data Temperature
LLL.L -15.6℃ or less LLL.L 4°F or less
023.5 23.5℃ 050.0 50°F
120.0 120.0℃ 250.0 250°F
HHH.H 160.6℃ or higher
HHH.H 321°F or higher