32
16. Cleaning the Prism
WARNING
Before running hazardous substance(s) through any system, necessary
precautions should be taken to ensure the safe handling of the
hazardous substance(s). If using a sample inlet unit, use caution when
disconnecting the instrument unit.
CAUTION
Cleaning liquids up to 160℃ can be used for CIP or SIP.
The momentary difference between the sample liquid temperature and
the cleaning liquid temperature must be no more than 80℃.
When using cleaning liquids at temperatures over 160℃, the power
source (DC24V) must be turned off. If the AD-32 (AD-33 or AD-34) is
used, unplug from the outlet before cleaning.
Detach the clamp band that
connects the main unit to the sample
inlet unit, piping or tank.
Clean the prism surface carefully
with a soft tissue soaked with warm
water or ethyl alcohol. If the sample
solution contains oil or grease, use
ethyl alcohol to ensure the prism
surface does not develop a film.
Development of a film on the prism
could cause erroneous
measurements.
NEVER clean the prism with an
abrasive material.
Cleaning the prism with an abrasive
material could cause scratches on
the prism which could lead to
erroneous measurements.
After cleaning is complete, re-attach
the instrument unit to the sample
inlet unit, piping or tank.
The procedure is described on page
14 of this instruction manual.
Fig. 16-1
Lukewarm
water
Ethyl
alcohol