6.0 Maintenance and Service
Changing the bacterial lter
Exchange daily or when blocked.
Check the bacterial lter
Set the vacuum regulator (
13
,g.1,page8)to
"maximum" (right stop).
As soon as the vacuum gauge shows a vacuum value
> -0.3 bar, while the suction hose is open,thelterhasto
be replaced.
Changing the bacterial lter
PleasedoonlyuseoriginalATMOSbacteriallters.The
devicemayneverbeoperatedwithoutDDSbacteriallter/
oversuction stop.
Bacteriallterplate/integratedbacteriallter:
Detach all hose connections carefully on the lid system
and take secretion canister out carefully to prevent spills
and contamination of the area. Dispose of secretions
properly.
Griplidsystemrmly,openlidoflterhousingbyturning
inanti-clockwisedirectionandremovelter.Disposeof
lterandthoroughlyrinseallpartsunderrunningwater.A
detergent or cleaning agent may also be used if required.
Aftercleaninganewbacterialltermustbeinserted.With
bacteriallterplates:smoothsidedown.
Externalbacteriallter:
Toopenthebacterialltercoverpleaseturnit.
Disposeofthebacteriallter.
Insertanewbacteriallterandclosecover.
) Make sure to clean the electrodes of the automatic
secretion canister evacuation!
) Mistake in polarity not possible!
) If no vacuum is achieved after switching on again the
suctionsystem,checkfreemovementoftheoat!
Fig. 34. External bacterial lter
Fig. 32. Lid with bacterial lter plate
Fig. 33. Lid with integrated bacterial lter
open