EasyManua.ls Logo

Atmos C 21 - 11.0 Hygiene Plan

Atmos C 21
44 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
39
11.0 Hygiene Plan
What How
Recommendations
When Who
Reusable parts
C
Cleaning
D
Disinfection
S
Sterilisation
After each
procedure
Weekly
Monthly
Qualifi ed and trained
staff who are familiar
with reprocessing.
(Please fi ll in the
responsible person
-> use a water-based
overhead marker)
Daily
Secretion canister
Housing for external
bacterial fi lter
X X
2)4)5)
Manual or automatic cleaning and disinfection X
/
Bacterial fi lter cover X X
2)4)5)
Manual or automatic cleaning and disinfection X
/
Bacterial fi lter plate with
blue marking /
bacterial fi lter
Disposable. Exchange daily or when blocked X
/
Lid / sealing rings X X
2)4)5)
Manual or automatic cleaning and disinfection X
Gasket X X
2)4)5)
Manual or automatic cleaning and disinfection X
Sleeve of overfl ow safety X X
2)4)5)
Manual or automatic cleaning and disinfection X
Float, closed side must up X X Manual or automatic cleaning and disinfection X
Secretion collection
jar
X X
2)4)5)
Empty when the jar is full; at least daily;
manual or automatic cleaning and disinfection
X
Disposable jar
system
Exchange and disposal of full jar X
Hose rinsing system
Suction nozzle
for hose rinsing
X X
3)
Wipe cleaning and disinfection X
Silicone attachment piece
X X
2)4)5)6)
Manual or automatic cleaning and disinfection X
Exchange of Silicone attachment X
Suction nipple with
suction hose (reusable)
X
Manual cleaning and disinfection after each patient;
rinse with hose disinfectant
X
X X
2)4)5)6)
Manual or automatic cleaning and disinfection X
Secretion suction hose
(disposable)
X
Rinse the secretion hose with the hose irrigation system
after each procedure
X
X
2)4)5)6)
Exchange or disinfection monthly X
Storage container for hose
rinsing system
X X
2)4)5)6)
Cleaning with a brush; Manual or automatic cleaning and
disinfection
X
Ear irrigation
Ear irrigation bowl X X
2)4)5)
Manual or automatic cleaning and disinfection X
Handle X X
3)
Wipe cleaning and wipe disinfection X
Jet connection X X
2)4)5)6)
Manual or automatic cleaning and disinfection X
Splash protection X
X
2)4)5)
Manual or automatic cleaning and disinfection X
Hose tip (disposable) Exchange after each application X
Rinsing lid with rinsing
hose
X X
2),4),5)
Manual or automatic cleaning and disinfection X
Rinsing bottle
X X
2),4),5),6)
Manual or automatic cleaning and disinfection; cleaning
in the dishwasher with the glass care programme
X
Ear irrigation system
(ATMOS
®
C 31)
X X
3)
Empty, dry, wipe, disinfect X
X
Rinse weekly with Bilpron (REF 510.2049.0) X
Medication nebulisation / Politzer
Handle for
compressed air
X X
3)
Manual cleaning and disinfection X
Sprayer jet
X Clean after every procedure X
X
2)4)5)6)
Manual or automatic cleaning and disinfection X
Sprayer head X X
2)4)5)
Multiple rinsing of the sprayer head with water X
Hose at sprayer head X X
Weekly exchange of the hose or when changing the
medication
X
Sprayer bottle X X
2)4)5)6)
Cleaning and disinfection (manual or automatic)
weekly or when changing the medication
X
Politzer olive X X
2)4)5)6)
Exchange after each application;
afterwards cleaning and disinfection
X
Politzer connection X X
2)4)5)6)
Exchange after each application;
afterwards cleaning and disinfection
X
Cleaning and disinfection plan