31
7.0 Troubleshooting
7.4 Suction system
7.5 Light sources
Error indication Possible cause Remedy
Weak suction or no suction at all • Suction hose is clogged • Rinse the suction hose with water (hose
can also be removed)
• Floatofoverowsafetyclosesthe
suction opening
• Checkllinglevelinthesecretioncanister
• Lid of secretion canister is not
closed tightly
• Check that lid of the secretion canister is
closed properly
• Bacteriallterisclogged • Replacethebacteriallter
• Hose connections are leaking • Seal up or replace connections
• Connecting hoses in the secretion
canister compartment are broken
• Check hose connections, remove kinks
• Secretion penetrated the suction
pump
• Have suction pump cleaned by a service
technician
No suction, but vacuum gauge
indicates -0.7 bar
• Suction hose is clogged • Rinse the suction hose with water (hose
can also be removed)
• Floatofoverowsafetyclosesthe
suction opening
• Checkllinglevelinthesecretioncanister
• Bacteriallterisclogged • Replacethebacteriallter
• Connecting hoses in the secretion
canister compartment are broken
• Check hose connections, remove kinks
• Overowsafetyisclosed • Hang up suction attachment, open
regulating valve and tap the secretion
canister;theoatsoftheoverowsafety
must drop back in their resting position
No suction and suction motor does
not start
• Photoelectric barrier / electronics
defective or photoelectric barrier is
polluted
• Clean photoelectric barrier
• Have photoelectric barrier checked by a
service technician
No suction; suction motor fails to start
although compressed-air pump starts
• Compressed-air handle and
suction hose in handle support
transposed
• Position handles corresponding to
switching function
Error indication Possible cause Remedy
No light • No voltage • Check mains voltage and fuses
• Light guide was swapped so
that another light channel was
activated.
• Position light guide correctly