1.2 Intended use
1.2.1 ATMOS C 21
Name: ATMOS C 21
Main functions:
Suction
Compressed air system for medication spraying and the
Politzer manoeuvre
Ear irrigation
Electric power supply for LED light handles and LED
headlight
Light sources
Storage and heating of endoscopes
Mirror quick heater
Instrument deposit and heating
Medical indications / application:
Standard ENT examination and/or therapy
Specication of the main function:
Suction at 40 l/min / -76 kPa
Alternatively suction at 55 l/min / -98 kPa
Compressed air for medication spraying, max. 2 bar
Compressed air for the Politzer manoeuvre, regulated
Ear irrigation with 37 °C ± 1 °C, max. 500 ml/min
Electric power supply for LED, 700 mA
Light source LED for light guide
Application organ:
Mouth to pharynx, auditory canal to the ear drum and the
nasal cavities
Application time:
ENT unit: Short term use (up to 30 days)
Suction / Compressed air / ear rinsing / light source:
Temporary application on the patient (less than 60
minutes)
Application site:
Application sites are clinics and practices for ENT doctors
and phoniatrists. The examination and/or therapy with
the ENT unit may only be executed by medically trained
persons.
Contraindications:
May not be used for irrigation of the paranasal sinuses. The
ear irrigation should not be applied to an infected auditory
canal or a perforated eardrum.
The product is: active
Sterility:
The ENT unit is no sterile product.
Single-use product / reprocessing:
The ENT unit is intended for multiple use. The device and
parts of the accessories are reusable. For information on
reprocessing, cleaning and disinfection, please see the
operating instructions.
1.2.2 ATMOS C 31
Name: ATMOS C 31
Main functions:
Suction
Compressed air system for medication spraying and the
Politzer manoeuvre
Ear irrigation
Electric power supply for LED light handles and LED
headlight
Light sources
Storage and heating of endoscopes
Mirror quick heater
Instrument deposit and heating
Medical indications / application:
Standard ENT examination and/or therapy
Specication of the main function:
Suction at 40 l/min / -76 kPa
Alternatively suction at 55 l/min / -98 kPa
Compressed air for medication spraying, max. 2 bar
Compressed air for the Politzer manoeuvre, regulated
Ear irrigation with 37 °C ± 1 °C, max. 500 ml/min
Electric power supply for LED, 700 mA
Light source LED for light guide
Application organ:
Mouth to pharynx, auditory canal to the ear drum and the
nasal cavities
Application time:
ENT unit: Short term use (up to 30 days)
Suction / Compressed air / ear rinsing / light source:
Temporary application on the patient (less than 60
minutes)
Application site:
Application sites are clinics and practices for ENT doctors
and phoniatrists. The examination and/or therapy with
the ENT unit may only be executed by medically trained
persons.
Contraindications:
May not be used for irrigation of the paranasal sinuses. The
ear irrigation should not be applied to an infected auditory
canal or a perforated eardrum.
The product is: active
Sterility:
The ENT unit is no sterile product.
Single-use product / reprocessing:
The ENT unit is intended for multiple use. The device and
parts of the accessories are reusable. For information on
reprocessing, cleaning and disinfection, please see the
operating instructions.