The ATMOS C 21 / C 31 may not be operated with units
not complying with the requirements of standard EN
60601-1 "Medical Electrical Equipment" and EN 60601-
1-2 "Electromagnetic compatibility (Medical Electrical
Equipment)".
The warranty will be rendered invalid in case of
damages caused due to the utilization of accessories or
consumables which are not approved by ATMOS for use
with this unit.
ATMOS is not liable for personal injury and damage to
property if
- no original ATMOS parts are being used,
- the advice for use in these operating instructions is
not being observed,
- assembly, new settings, alterations, extensions
and repairs have been carried out by personnel not
authorised by ATMOS.
Please pay also attention to the safety information in
following chapters.
Please note:
A medical insulating transformer with earth leakage
monitor or any similar safety system acc. to EN 60 601-1
is required, if several devices are connected over one
common power supply. The transformer must correspond
to the power consumption of all the devices to be
connected.
When connecting several devices on one grounding
receptacle, the allowed strain and leakage current have
to be observed!
Never leave the patient unattended at the treatment unit.
Itisnotallowedtouseammablesubstanceswiththe
device.
2.0 For your safety