USER MANUAL 29
__________________________________________________________________________________________________________________
Calibration with external weight:
Press MENU key.
Press
T
key when CALIb function
appears.
The following options will be displayed:
-CAL on – calibration with external
recommended standard of mass (see
technical data).
-CAL StP – calibration with external weight,
confirmation of successive steps - MENU key,
out – leave without changes
- CAL Prn – calibration report,
- CAL tM – set time interval for internal
calibration,
- CAL
C – set temperature difference for
internal calibration,
- out.
Press
T
key when CAL StP option
appears (calibration in two steps).
Press
T
key when weight value used for
calibration is indicating or use othEr option
and inscribe proper value ( keys
0
,
,
T
)
Press MENU and wait for writing zero to the
scale.
When LOAD message appears put standard
of mass on the pan. Press MENU key (CAL on
doesn’t need pressing MENU key).
Wait until internal calibration is finished and
zero indication is displayed.