Page 46 of 59 BACTcontrol – User Manual – V0.4.4 microLAN b.v.
Remove stirrer
Remove the stirrer with the stirrer removal tool
Be careful! Stirrer is very small
Cleaning reaction chamber and stirrer
Check for possible biofilm or pollution inside the
reaction chamber.
Clean the measurement cell, the glass of the lens
and the stirrer with the cleaning solution and dry
with a paper cloth
Reinstall the stirrer and filter and closing reaction chamber
Place the stirrer in the reaction chamber.
Place the ceramic filter with the C-ring, make
sure that it is installed with the upside on top.
The upside is marked with a small dot as shown
in the figure on the right. Close reaction
chamber and turn the knob hand-tight.
Make sure that cover is placed properly to
prevent any leakage.