Sensor positions
threshold exactly at
the presented condition
Threshold position
adjusted by
Manual Adjust
Darkest
(no signal)
Most Light
(saturated
signal)
Output OFF Output ON
Condition
Presented
Figure 12. Calibraon SET (Light Operate shown)
Calibraon SET and Manual Adjust
Moves switching threshold value up or down to make adjustments
•
Slide Mode switch to ADJ to enter Adjust mode
• Press "+" to increase; press "-" to decrease
◦ GREEN display shows the switching threshold value
◦ 2 seconds aer adjustment, the GREEN display will ash 3 mes to conrm
• Slide Mode switch to RUN to complete operaon
Remember: Auto Thresholding is automacally disabled in Calibraon SET
Follow these steps to perform a Calibraon SET:
Note: TEACH Selecon must be programmed to CAL SEt.
1. Enter Adjust Mode
Method Acon Result
SET Buon
20
• Set Mode switch to ADJ Display: Red - Signal Level; Green -
Threshold
Remote Input
21
No acon required; sensor is ready for
Calibraon SET method
2. SET Sensing Condion
Method Acon Result
SET Buon • Present sensing condion
• Click the SET rocker buon
Threshold Condion Accepted Displays
read "cAL SEt" then ashes "PASS"; Sensor
returns to Adjust mode
Threshold Condion Unacceptable
Displays read "cAL SEt" then ashes "FAIL";
Sensor returns to Adjust mode
Remote Input • Present sensing condion
• Single-pulse the remote input
3. Return to RUN Mode
20
SET Buon: 0.04 seconds ≤ "Click" ≤ 0.8 seconds
21
Remote Input: 0.04 seconds ≤ T ≤ 0.8 seconds
DF-G3 Long Range Expert
™
Dual Display Fiber Amplier
24 www.bannerengineering.com - Tel: 763.544.3164