5. Advanced operation
5.12 Amb ient Light Compensation (ALC)
About ALC
Ambient Light Compensation (ALC) can only be enabled on your display w hen the DI-
COM display function is selected. Therefore, plea se refer to "Display functions", page
24 to correctly set the display function.
When ALC is enabled, the DICOM display function will be recalculated taking a preset ambient light cor-
rection value into account. This value is determined by the selected reading room. Therefore, it is also
important to select a realistic reading room when enabling ALC. This can be done by following the ins
truc-
tions in "Reading rooms", page 26.
To enable/disable ALC:
1. Bring up the OSD main menu.
2. Navigate to the Conguration > Calibration > Ambient Light menu.
3. Enter the Ambient Light Compensation submenu.
4. Select Enabled/Disabled as desired and conrm.
5.13 Reading roo ms
About reading rooms
Reading rooms can only be selected when the DICOM display function is selected.
Therefore, please refer to "Display functions", page 24 to correctly set
the display
function.
The American Association of Physicists in Medicine (AAPM) composed a list of pre-dened reading rooms.
Each of these reading rooms are dened by following param
eters:
the maximum light allowed in this type of room
the preset ambient light correction value for this reading room
These parameters are stored in your display and determine the preset ambient light correction value to
take into account to recalculate the DICOM display function when Ambient Light Compensation (ALC) is
enabled. Please refer to "Ambient Light Compensation (ALC)", page 26 to enable ALC.
The available reading rooms for your Nio 3MP LED Display are:
CR/DR/ MAMMO: Corresponds to light conditions in diagnostic reading rooms for computed radiology,
digital radiology or mammography. This setting has the lowest maximum ambient light.
CT/MR/NM: Corresponds to light conditions in diagnostic reading rooms for computed tomography,
magnetic resonance or nuclear medicine scans.
Staff Ofce: Corresponds to light conditions in ofce rooms.
Clinical Viewing Room: Corresponds to light conditions in diagnostic reading rooms for clinical view-
ing.
Emergency Room: Corresponds to light conditions in emergency rooms.
Operating Room: Corresponds to light conditions in operating rooms. This setting has the highest
maximum ambient light.
To select a reading roo m:
1. Bring up the OSD main menu.
26
K5902114 NIO 3MP LED DISPLAY 03/10/2016

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Barco MDNC-3421 and is the answer not in the manual?

Barco MDNC-3421 Specifications

General IconGeneral
BrandBarco
ModelMDNC-3421
CategoryMonitor
LanguageEnglish

Related product manuals