8. Important information
8. IMPORTANT INFORMATION
8.1 Safety information
General recommendations
Read the safety and operating instructions before operating the device.
Retain safety and operating instructions for future reference.
Adhere to all warnings on the device and in the operating instructions manual.
Follow all instructions for operation and use.
Electrical Shock or Fire Hazard
To prevent electric shock or re hazard, do not remove cover.
No serviceable parts inside. Refer servicing to qualied personnel.
Do not expose this apparatus to rain or moisture.
Modications to the unit:
Do not modify this equipment without authorization of the manufacturer.
Type of protection ( electrical):
Display with external power supply: Class I equipment.
Degree of safety (ammable a nesthetic mixture):
Equipment not suitable for use in the presence of a ammable anesthetic mixture
with air or with oxygen
or nitrous oxide.
Non-patient care equipment
Equipment primarily for use in a health care facility that is i
ntended for use where contact with a patient
is unlikely (no applied part).
The equipment may not be used with life support equipment.
The user should not touch the equipment, nor its signal input por
ts (SIP)/signal output ports (SOP)
and the patient at the same time.
Power connection Equipment with external 24 VDC power suppl
y
Power requirements: The equipment must be powered using the d
elivered medica l approved 24 VDC
(
) SELV power supply.
The medical approved DC (
) power supply must be powered by the AC mains voltage.
The power supply is specied as a part of the ME equipme
nt or combination is specied as a ME
system.
To avoid the risk of electric shock, this equipment must only be connected to a supply mains with
protective earth.
The equipment should be installed near an easily accessible outlet.
The equipment is intended for continuous operation.
Transient over-voltage
If the device is not used for a long time, disconnect it from the AC inlet to avoid damage by transient
over-voltage.
K5902114 NIO 3MP LED DISPLAY 03/10/2016