EasyManua.ls Logo

Baur Sys compact 2000 M pro - Performing the TDR Measurement

Baur Sys compact 2000 M pro
98 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Syscompact 2000 M pro Cable fault pre-location
822-175-2 55 / 98
Pre-location
parameters
Settings Function
Pulse width
20 ns 10 µs Is used to set the pulse width
The pulse width is set automatically depending on the measur
ing
range. Narrow pulses have high resolution, however, they have a
lower range due to the higher damping. They are especially
suitable for shorter cables. When using wide pulses, the range
increases and the resolution decreases.
Voltage
10 60 V Is used to set the pulse voltage
Impedance
10 250 Ohm Is used to set the output impedance
Automatic cursors
On/Off Enables or disables the automatic placement of cursors
When On is selected, a cursor is placed automatically after a
measurement if the cable end or a cable fault is detected.
Voltage proof
(Can only be set with TDR
methods TDR and TDR
Continuous)
On/Off Enables and disables the input protection of the time domain
reflectometer
If On
is selected, the input protection is activated up to AC 400 V;
measurements are possible on live cables up to AC 400 V.
Before each measurement on live cables, check whether
the fuses on the terminals of the TDR connection cable are
working.
If the TDR connection cable is used without a fuse or with
another type of fuse, the IRG 2000 must not be connected
to a live cable.
Trigger current level
(Can only be set with TDR
method TDR Trigger)
10 1000 A Is used to set a threshold value for current that is
detected by an
external measuring device connected to the Trigger Input
port. If
this threshold value is exceeded, the IRG 2000 performs a
measurement.
12.3.4 Performing the TDR measurement
Prerequisites
The IRG 2000 is used as an independent device.
Further information: Chapter Using the IRG 2000 as an independent device (on page 42)
The TDR connection cable is connected to the IRG 2000 and the test object correctly.
Further information: Chapter Connecting the TDR connection cable to the test object (IRG
2000) (on page 37)
For measurements on live cables: Chapter Safety instructions for connecting the IRG 2000
to the live cable (on page 37)
The general pre-location parameters and the TDR parameters have been set.
Further information: Chapter Setting the general pre-location parameters on the IRG 2000
(on page 52), chapter Setting TDR parameters (on page 54)

Table of Contents