BD PrepMate™ Instrument User’s Manual
26
HORIZONTAL IN FAULT
Troubleshooting
Turn off the instrument. Pull out tray; turn on the instrument again.
HORIZONTAL OUT FAULT
Troubleshooting
Turn off the instrument. Push in tray; turn on the instrument again.
OTHER FAULTS
Should any of the following faults occur or messages appear on the screen, call
Becton,
Dickinson and Company Technical Support.
Horizontal Position Fault
Vertical Top Fault
Vertical Mid Fault
Vertical Bottom Fault
Plunger Driver Fault
Rear Travel Limit
Draw Sensor Fault
Draw Sensor or Connection Failure
Plunger Sensor Fault
Tray accidentally moved to the rear travel position
Fiber optic sensor heads misaligned
Vial old-down door hinge screws loose
NOTE
If there is a loss of power to the BD PrepMate™ Automated
Acc
essory, the processor does not retain memory. The operator will
need to determine which step the process was in and restart testing
at the appropriate step.