BD PrepMate™ Instrument User’s Manual
28
12 Operational Specifications
12.1 Power requirements
Rated Voltage: 100–240 VAC, 50–60 Hz
Maximum Current: 2.5 A
12.2 Operating conditions
Ambient Temperature: 18–30°C (64–86°F)
Ambient Humidity: 20–80% RH (non-condensing)
12.3 Storage conditions
Temperature: 0–50 °C, 32–122 °F
12.4 General
Weight: 80 lb, 36.3 kg
Height: 22”
Width: 14”
Depth: 23”
12.5 Fusing
The BD PrepMate™ has 2 fuses, located in the power entry module. All fuses are 5x20 mm and
are 3.15 A @ 250 VAC time-delay, IEC 60127 rated. The product name plate is located on the
rear panel with the power inlet and provides fuse information. Replacement fuses must be the
same rating and type (T) as identified. Type T indicates a time-delay acting fuse.