Transducer
A green LED inside the transducers transparent housing indicates whether the
transducer is functioning properly (see table 2-4 on page 2-11).
Table 2-4. Transducer LED Troubleshooting Codes
Error Codes Possible Cause Corrective Action
Slow steady flash Transducer is functioning
properly.
No action is required.
Rapid flash Device error. Replace the transducer.
LED off No power to the transducer,
or the transducer has failed.
Connect the power, or
replace the
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728) Page 2-11
Error Codes and Troubleshooting Tips
Chapter 2: Troubleshooting Procedures
2