Detailed Component Descriptions
The MedPlus TotalAlert® Alarm Network can consist of the following: Master
Alarm, Area Alarm, Local Alarm, Area Alarm Network Communications Board,
and gas-specific transducers that include the appropriate DISS, gas-specific,
demand valves.
Master Alarm
The Master Alarm continuously monitors up to 60 input signals from source
equipment by using dry contact switches (see figure 1-11 on page 1–21). The
monitored signals are supervised for open-circuit conditions.
Figure 1-11. Master Alarm
NOTE:
The available 60-Point Master Alarm is two 30-Point Master Alarms mounted in
one housing.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728) Page 1-21
Detailed Component Descriptions
Chapter 1: Introduction