EasyManua.ls Logo

Binder APT.line C 150 - Operating Mode USER: Advanced Settings

Binder APT.line C 150
79 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
C 150 (E2) 06/2011 page 31/78
7.3 Operating mode USER: Advanced settings
• In Normal display, press “MODE” and “EXIT” simultaneously for 3 seconds to access the options for
selecting the C 150's operating modes.
Figure 21: Selecting the operating mode
• Press “▲”until the value “OP.U” (operating mode USER) appears in the upper display.
• Press “MODE”. The display to enter the password appears.
The operating mode USER is password-protected by a number. The password is preset to "1" in factory.
You can change the password in operating function “PA.U”.
Figure 22: Password request
• Use “▲” or “▼” to enter the password and press “MODE”.
• Use “MODE” to successively call up the operating functions.
• Use “▲” or “▼”to set the values of the operating functions.