EasyManua.ls Logo

Binder APT.line C 150 - CO2 Sensor Disinfection; Hot-Air Sterilization Cycle

Binder APT.line C 150
79 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
C 150 (E2) 06/2011 page 59/78
15.3 Disinfection of the CO
2
sensor
To ensure complete disinfection and proper function of the sensor, BINDER recommends a wipe disinfec-
tion of the sensor head with pure alcohol or non-corrosive alcohol-based surface disinfectants. The disin-
fectant must be non-corrosive and free of chlorine or any acid. We recommend using the disinfectant Art.
No. 1002-0022. Avoid strong shocks when handling the CO
2
sensor.
CAUTION
Excess temperature.
Immersion of sensor into liquids.
Shocks of the CO
2
sensor.
Damage to the CO
2
sensor.
Do NOT immerse the CO
2
sensor into liquids.
Do NOT expose the CO
2
sensor to autoclaving.
Do NOT expose the CO
2
sensor to hot-air sterilization.
Avoid strong shocks of the CO
2
sensor (by putting it down hard, or dropping).
We recommend regular disinfection of the CO
2
sensor.
CAUTION
Connecting or removing the CO
2
sensor during operation.
Damage to the CO
2
sensor.
Connect or remove the CO
2
sensor only with the unit turned off.
Recommended procedure:
1. Turn off the unit
2. Pull out the sensor
3. Spray the sensor head with alcohol or wipe it clean with a soaked cloth.
The reaction time of the disinfectant used must be followed correctly
4. Before reinserting the CO
2
sensor, it must be completely dry.
5. The filter in the front of the sensor only needs replacing when damaged or dirty.
The CO
2
sensor head was especially adjusted for the specific chamber. To avoid confusion, an adhesive
label with a serial number is adhered to the sensor head. When exchanging the sensor, repeat CO
2
ad-
justment.
CAUTION
Different CO
2
sensor.
Invalid calibration.
Do NOT change the CO
2
sensor head.
¾ Write down the serial number of the CO
2
sensor.
15.4 Hot-air sterilization at 180 °C / 356 °F
The C 150 has an automatically controlled sterilization cycle that sterilizes at 190 °C / 374 °F. It consists
of the following steps:
Heating up phase: The interior is rapidly heated to 180 °C / 356°F

Table of Contents

Related product manuals