C 150 (E2) 06/2011 page 60/78
Sterilization phase: The temperature is maintained for 3 hours, which ensures that 180 °C / 356 °F is
reached on all internal surfaces for at least 30 minutes.
A cooling-down phase follows until 37 °C / 98.6 °F is reached.
The precise duration of the entire sterilization cycle depends on the ambient temperature at the installa-
tion site and can thus vary. At an ambient temperature of 25 °C / 77 °F, the total duration is approx. 12
hours. The display "End" in the lower controller display (Figure 35 page 62) indicates the end of the ster-
ilization procedure. You can shorten the cooling-down time by aborting the sterilization cycle (chap. 15.5)
during the cooling-down phase, i.e. no sooner than after 6 hours.
During sterilization, the CO
2
valve is closed and the CO
2
controller turns off entirely.
The CO
2
sensor is temperature resistant up to a maximum temperature of 60 °C / 140 °F.
CAUTION
Excess temperature.
Damage to the CO
2
sensor.
Do NOT expose the CO
2
sensor to hot-air sterilization.
The CO
2
sensor head was especially adjusted for your specific chamber. To avoid confusion, an adhe-
sive label with a serial number is adhered to the sensor head. When exchanging the sensor, repeat CO
2
control adjustment.
CAUTION
Different CO
2
sensor.
Invalid calibration.
Do NOT change the CO
2
sensor head.
¾ Write to down the serial number of the CO
2
sensor.
The safety controller settings are inactive during sterilization. They become functional again
following abortion of the sterilization and/or restart of the unit at the main power switch.
Before carrying out the first hot-air sterilization, remove any protective lamination sheet from
the inner metal surfaces.
Procedure for hot-air sterilization:
Turn off the unit
Pull out the sensor
Empty the water pan.
WARNING
Danger of implosion.
Damage to the unit.
¾ Empty the water pan before starting the hot-air sterilization.
Clean the C 150.
Water pan and shelves must be inside the incubator, the water pan at its usual place on the bottom.
Close the inner glass door and the outer unit door.
Turn on the unit.