C 150 (E2) 06/2011 page 63/78
CAUTION
Glass door and inner chamber become hot during sterilization.
Danger of burning.
∅ Do NOT touch the glass door and inner surfaces for approx. 4 hours after aborting ster-
ilization.
Aborting sterilization after less than 6 hours
When aborting the sterilization prematurely, it may be that the cells/pathogens inside the unit have not all
been killed. You should repeat the sterilization.
CAUTION
Interruption of temperature reaction time.
Ineffective sterilization.
¾ Repeat the sterilization.
CAUTION
Glass door and inner chamber become hot during sterilization.
Danger of burning.
∅ Do NOT touch the glass door and inner surfaces for approx. 7 hours after aborting ster-
ilization.
• When the inner chamber has cooled down to a value below 60 °C / 140°F, turn off the unit and plug in
the CO
2
sensor.
The CO
2
sensor is temperature resistant up to a maximum temperature of 60 °C / 140°F.
• Turn on the unit (chap. 5).
Three ev
ents which can abort the automatic sterilization cycle prematurely:
• A manual abortion (chap. 15.5.1)
•
Opening the outer door (chap. 15.5.2)
•
Turning off the power to the C 150 or a brief power failure (chap. 15.5.3)
15.5.1 Aborting the hot-air sterilization manually
You can abort the sterilization cycle manually.
• Press the controller button “180 °C” for three seconds during the sterilization cycle.
The lower display will show the operating function “StE”, and the value “ON” will appear in the upper
display.
• Press “▲”.
The upper display toggles from “ON” to “OFF“.
• Press “MODE” to end the sterilization cycle.
• Do not open the doors of the C 150 until the interior temperature has dropped to 37 °C / 98.6°F.