BD / BF / ED / FD / FED (E3.1) 12/2017 page 27/105
6.6.1 Set-point entry for temperature in two-door chambers (ED, FD, FED 720)
With two-door chambers (ED, FD, FED 720) the maximum adjustable temperature set-point depends on
the ambient temperature:
Range from 18 °C up to 26 °C: maximum temperature set-point: 300 °C
Range from above 26 °C up to 40 °C: maximum temperature set-point decreasing with increasing
ambient temperature
Temperature set-point [ °C ]
Ambient temperature [ °C ]
Figure 9: Maximum temperature set-point depending on the ambient temperature
Ambient temperature Maximum temperature set-point
18 °C bis 26 °C 300 °C
28 °C 298 °C
30 °C 296 °C
32 °C 294 °C
34 °C 292 °C
36 °C 290 °C
38 °C 288 °C
40 °C 286 °C
This ensures the maximum lifetime of the controller.