CB-S / CB-S-UL (E7) 06/2020 page 105/119
The technical data refers to the so defined usable volume.
Do NOT place samples outside this usable volume.
Do NOT load this volume by more than half to enable sufficient airflow inside the CO
2
incuba-
tor.
Do NOT divide the usable volume into separate parts with large area samples.
Do NOT place samples too close to each other in order to permit circulation between them
and thus obtain a homogenous distribution of temperature and CO
2
.
26.4 CB-S / CB-S-UL technical data
Depth including door handle, I-triangle, and
connections
mm / inch 795 / 31.30 855 / 33.66
Wall clearance rear (minimum)
Wall clearance side (minimum)
Number of inner glass doors
Number of shelves, series
Size of shelf, width x depth
Maximum permitted total load
Temperature range, 6 °C / 10.8 °F above ambient,
up to
°C / °F 50 / 122 50 / 122
Temperature uniformity (variation) at 37 °C/ 98.6
°F
+/- K 0.3 0.4
Recovery time after door open for 30 sec
at 37 °C / 98.6°F
minutes 8 9