Why is the CO2 concentration inside my Binder CB-S 170 Accessories chamber too high?
- MMichael DavisSep 7, 2025
If the CO2 concentration inside the chamber is too high, calibrate and adjust the controller.
Why is the CO2 concentration inside my Binder CB-S 170 Accessories chamber too high?
If the CO2 concentration inside the chamber is too high, calibrate and adjust the controller.
Why is the temperature inside my Binder Accessories chamber too low?
If the temperature inside your Binder Accessories chamber is too low, make sure the doors are properly closed. If they are, inspect the door gaskets and replace them if they're defective. Also, calibrate and adjust the controller.
What to do if the temperature inside the Binder Accessories chamber is too high?
If the temperature inside your Binder Accessories chamber is too high, consider these possible causes: the installation site may be too warm, the difference between the set temperature and the ambient temperature may be too low, or there may be too much external heat load. To resolve this, select a cooler place for installation, ensure at least a 6 °C difference between the set and ambient temperatures, and reduce any external heat load. If sterilization terminated prematurely, let the chamber cool down and calibrate and adjust the controller.
What does it mean when the Binder Accessories alarm message “Door open” appears?
The alarm message “Door open” indicates that the outer chamber door is open. Ensure that you close the chamber door properly to resolve this.
What does the alarm message “Safety controller” on my Binder CB-S 170 mean?
If you're seeing the alarm message “Safety controller” on your Binder Accessories, it could be because the safety controller is set too low. Check the setting of the safety controller and select a suitable value if needed. If sterilization terminated prematurely, let the chamber cool down.
Why are the set-point values not equilibrated in my Binder CB-S 170 Accessories?
If the set-point values are not equilibrating, the operating mode “Idle mode” may be activated. Deactivate the operating mode “Idle mode”.
What to do if there is no chamber function and a dark display on my Binder CB-S 170 Accessories?
If there is no chamber function and the display is dark, check if the main power switch is off and turn it on.
What does it mean when the Binder CB-S 170 Accessories display shows “ – – – – ” and the alarm message is “CO2 sensor defective”?
If the display shows “ – – – – ” and the alarm message is “CO2 sensor defective”, it means the CO sensor is defective. Turn off the chamber.
Why is the CO2 concentration inside my Binder Accessories chamber too low, and why is the alarm message “CO2 range” appearing?
If the CO2 concentration is too low and you're seeing the alarm message “CO2 range”, it means the current actual CO value is outside the tolerance range. Check the tolerance range settings. Remove the respective cause if there are other error messages.
What does the Binder CB-S 170 Accessories alarm message “Door heating sensor” mean?
If you see the alarm message “Door heating sensor”, it means the door heating sensor is defective.
Brand | Binder |
---|---|
Model | CB-S 170 |
Category | Accessories |
Language | English |
Defines the necessary qualifications for personnel installing, testing, and operating the chamber.
Instructions on handling and using the operating manual for safe operation.
Provides a visual overview of the CO2 incubator CB-S / CB-S-UL.
Details the controller display and USB interface on the instrument panel.
Procedure for unpacking and verifying the chamber and its accessories.
Specifies requirements for the chamber's installation site and environment.
Instructions for connecting and general notes regarding the CO2 sensor.
Details on connecting the CO2 gas supply to the chamber.
Guidance on connecting the chamber to the power supply.
Explains the controller's menu layout and user authorization levels.
Step-by-step instructions for powering on the chamber.
Lists the default settings provided by the manufacturer.
Procedure to set altitude for accurate CO2 concentration calculation.
Instructions for setting the desired temperature value.
Instructions for setting the desired CO2 concentration value.
Guide to setting and changing User and Admin passwords.
Details on the electronic overtemperature safety controller.
Procedure to change the language of the controller interface.
Instructions for setting the chamber's current date.
Instructions for setting the chamber's current time.
Defining the acceptable deviation for temperature alarms.
Defining the acceptable deviation for CO2 alarms.
Overview of alarm messages and their causes.
Actions to take for specific alarm conditions.
Accessing and displaying current network configuration details.
Procedure for modifying IP address, DNS, gateway, and subnet mask.
Description of the data stored by the chamber's data recorder.
Configuring the interval for saving recorder data.
How to load configuration data from a USB stick.
How to save configuration, recorder, and service data to USB.
Methods for verifying CO2 concentration accuracy.
Methods for verifying temperature measurement accuracy.
Guidelines for personnel practices to prevent contamination.
General recommendations for handling the incubator to prevent contamination.
Detailed procedures for cleaning the chamber and its components.
Methods for chemical decontamination and disinfection of the chamber.
Specific instructions for disinfecting the CO2 sensor.
General description and phases of the hot-air sterilization cycle.
Steps and preparations required for conducting hot-air sterilization.
Procedure for initiating and managing the sterilization cycle.
Information on the consequences of stopping the sterilization early.
Requirements for personnel involved in maintenance and service.
Information on regular maintenance schedules and service.
Basic troubleshooting steps for common chamber issues.
Guidance on how to dispose of packaging materials.
Steps for safely taking the chamber out of operation.
Regulations for disposing of the chamber in Germany.
Information on the chamber's factory calibration procedures.
Provides detailed technical specifications for the chamber models.
Lists standard equipment and available optional accessories.
Official declaration confirming compliance with EU directives.
Form for declaring safety and health information for international returns.