CB-S / CB-S-UL (E7) 06/2020 page 78/119
A warning sticker is located above each access port.
When operating a chamber with silicon access ports, both silicon plugs must tightly close the access
ports. If the plugs are inserted in a not-gastight manner, or if plugs are missing, CO
2
gas may escape into
the environment. The CO
2
control only turns off when the chamber door is opened.
DANGER
Danger of suffocation and poisoning by high concentration of CO
2
(> 4 Vol.-%).
Death by suffocation
Tightly close each access port with two plugs during operation.
20.2 Base on castors (option)
In order to obtain easy access to the incubator and to avoid contamination of the incubator caused by soil
pollution, BINDER recommends using the base on castors.
The mounting instructions 7001-0147 delivered with the base on castors describe its in-
stallation (Art. No. 9051-0024).
20.3 Flat stacking frame (option)
We recommend not stacking CO
2
incubators directly on top of one another in order to avoid transmission
of shocks and vibrations from one chamber to the other. This could happen e.g. while opening or closing
the door, cleaning, loading and unloading the chamber. BINDER offers a flat stacking frame for safe
stacking of two incubators.
20.4 APT-COM™ 4 Multi Management Software (option)
The chamber is regularly equipped with an Ethernet interface (15) that can connect the BINDER APT-
COM™ 4 Multi Management Software. The MAC Address is indicated in the “Ethernet” controller menu
(chap. 16.1.1). The actual temperature and CO
2
values are given at adjustable intervals. Programming
can be performed graphically via PC. Up to 100 chambers can be cross-linked. For further information
please refer to the APT-COM4 operating manual.
20.5 Data logger kit (option)
The BINDER Data Logger Kit offers an independent long-term measuring system for temperature. It is
equipped with a keyboard and a large LCD display, alarm functions and a real-time function. Measure-
ment data are recorded in the Data Logger and can be read out after the measurement via the RS232
interface of the Data Logger. It offers a programmable measuring interval and permits storing up to 64000
measuring values. Reading out is done with the Data Logger evaluation software. You can give out a
combined alarm and status protocol directly to a serial printer.
Data Logger Kit T 220: Temperature range -90 °C / -130 °F up to +220 °C / 428 °F
For detailed information on installation and operation of the BINDER Data Logger, please
refer to the mounting instructions Art. No. 7001-0204 and to the original user manual of the
manufacturer, supplied with the data logger.