CB-S / CB-S-UL (E7) 06/2020 page 106/119
2
2
2
2
Recovery time after door open for 30 sec
at 5 vol.-% CO
2
minutes 5 5
2
Connection hose nozzle DN6 for CO
2
Connection to the chamber for hose with internal
mm / inch 6 / 0.24 6 / 0.24
Electrical data (model versions C170-230V, CBS260-230V)
IP system of protection acc. to EN 60529
Nominal voltage
(+/-10%)
Power plug (IEC connector plug)
Installation category acc. to IEC 61010-1
Pollution degree acc. to IEC 61010-1
5x20mm / semi time-lag / 10 A
Different electrical data for C-UL constructed for the USA and Canada
(model versions C170UL-120V, CBS260UL-120V)
Nominal voltage (+/-10 %) at 50 Hz power frequen-
cy
Nominal voltage (+/-10 %) at 60 Hz power frequen-
cy
Power plug (IEC connector plug)
6.3 X 32 mm / 250V / super-time-lag TT / 16A
Environment-specific data
Energy consumption at 37 °C / 98.6°F
The recovery times of the gas concentrations inside the chamber following the door being opened coin-
cide with a connection pressure of 2.0 bar / 29 psi. Decreasing supply pressure results in longer recovery
times.
All technical data is specified for unloaded chambers with standard equipment at an ambient temperature
of +22 °C +/- 3 °C / 71.6 °F +/- 5.4 °F and a power supply voltage fluctuation of +/-10%. The technical
data is determined in accordance to BINDER Factory Standard Part 2:2015 and DIN 12880:2007.
All indications are average values, typical for chambers produced in series. We reserve the right
to change technical specifications at any time.
26.5 Important conversion data for non-SI units
1 ft = 0.305 m = 0.000305 km
1 m = 100 cm = 3.28 ft = 39.37 inch
1 km = 1000 m = 3280.83 ft
1 mbar = 0.0145 psi