CB-S / CB-S-UL (E7) 06/2020 page 83/119
22.1 Cleaning
Disconnect the CO
2
incubator from the power supply before cleaning. Disconnect the power plug.
The interior of the chamber must be kept clean. Thoroughly remove any residues of the charg-
ing material.
Wipe the surfaces with a moistened towel. In addition, you can use the following cleaning agents:
Exterior surfaces,
instrument panel:
Standard commercial cleaning detergents free from acid or halides.
Alcohol-based solutions.
We recommend using the neutral cleaning agent Art. No. 1002-0016.
Inner chamber,
shelves,
water pan:
Standard commercial cleaning detergents free from acid or halides.
Copper sulphate solutions or alcohol-based solutions.
We recommend using the neutral cleaning agent Art. No. 1002-0016.
CO
2
sensor
Alcohol-based solutions
Do not immerse the CO
2
sensor into the solution.
Disinfection with alcohol or an alcohol-based surface disinfectant without corro-
sive effect, free from acid or halides. We recommend using the disinfectant spray
Art. No. 1002-0022.
Silicone door gasket: Alcohol-based solutions or neutral cleaning agent Art. No. 1002-0016.
Zinc coated hinge
parts, rear chamber
wall
Standard commercial cleaning detergents free from acid or halides. Do NOT use
a neutral cleaning agent on zinc coated surfaces.
Do not use cleaning agents that may cause a hazard due to reaction with components of the device or
the charging material. If there is doubt regarding the suitability of cleaning products, please contact
BINDER service.
We recommend using the neutral cleaning agent Art. No. 1002-0016 for a thorough cleaning.
Any corrosive damage that may arise following use of other cleaning agents is excluded from
liability by BINDER GmbH.
Use only the products recommended by BINDER for cleaning and disinfection
NOTICE
Danger of corrosion by using unsuitable cleaners.
Damage to the chamber.
∅ Do NOT use acidic or chlorine cleaning detergents.
∅ Do NOT use a neutral cleaning agent on other kind of surfaces e.g., the zinc coated
hinge parts or the rear chamber wall.
For surface protection, perform cleaning as quickly as possible.
After cleaning completely, remove any cleaning agents from the surfaces with a moistened
towel. Let the chamber dry.
Soapsuds may contain chlorides and must therefore NOT be used for cleaning.
With every cleaning method, always use adequate personal safety controls.